NT8

®

Resources

Publications

Crystal structure of the chemokine receptor CXCR4 in complex with a viral chemokine
Qin L, Kufareva I, Holden LG, Wang C, Zheng Y, Zhao C, Fenalti G, Wu H, Han GW, Cherezov V, Abagyan R, Stevens RC, Handel TM.
Science. 2015 Mar 6;347(6226):1117-22. doi: 10.1126/science.1261064. Epub 2015 Jan 22. 
PMID: 25612609 

Computer Requirements
  • Windows XP or Newer (Windows Vista, Windows 8, Windows 10)
  • 1 open USB port (compatible with USB 3.0 or older)
  • 1 GB RAM (32-bit) or 2 GB RAM (64-bit)
  • 1 GB hard disk space
Electrical Specifications

110-240 V, 50-60 Hz, 130 W typical, 200 W max. Standard or European outlet

Physical Dimensions

Width: 653 mm (25.7")

Depth: 511 mm (20.1")

Height: 431 mm (16.9")

Weight: 38 kg (83.8 lb)

High Volume Pipette Tips

To order more high volume pipette tips for the Plate Copy Option (BrandTech PN 73102), contact BrandTech Scientific or your preferred lab products distributor.

Waste/Wash Source Bottle Recommendations

We recommend a 2 L or greater waste bottle for the wash station. It is recommended to use an even larger wash solution bottle since the wash solution is used as the system liquid.

Drop Dispensing Accuracy
 VolumeLiquidCV's
Wet Dispense100 nLWater2.67%
Dry Dispense100 nL 100% DMSO 3.5% 
Dry Dispense100 nL 10% Isopropanol 4.1% 
Dry Dispense100 nL Sticky Protein 2.7% 
Dry Dispense100 nL Easy Protein 4.3% 
Dry Dispense100 nL 50% PEG 6K 5.5% 
Dry Dispense100 nL Lysozyme 4.5% 
Dry Dispense100 nL 70% MPD 4.2%